  • Development of a liquid chromatography method for the analysis of josamycin.

Development of a liquid chromatography method for the analysis of josamycin.

Journal of pharmaceutical and biomedical analysis (2008-03-18)
F Daidone, R Heuvelmans, L Aerden, J Hoogmartens, E Adams
ABSTRACT

Out of three methods for the analysis of josamycin, the best one was selected and used as starting point for further development. A central composite design was applied to find the most influencing parameters and to optimize the chromatographic conditions and a full factorial design was used to perform a robustness study. The final method uses a Hypersil ODS column 5microm, 250mmx4.6mm i.d. maintained at 45 degrees C. The mobile phase is composed of acetonitrile-phosphate buffer (pH 3, 0.2moll(-1))-tetrabutylammonium hydrogen sulphate 0.2moll(-1)-water (21:5:3:71, v/v/v/v). Strongly retained impurities after the main peak require gradient elution, which is obtained by increasing linearly the acetonitrile concentration (from 21% to 50%, v/v) and decreasing the TBA concentration (from 3% to 0%, v/v) in the mobile phase. The total run time was 65min. UV detection is performed at 232nm and the flow rate is 1ml/min. The method shows good selectivity, precision, linearity and sensitivity. Five commercial bulk samples were analyzed.

MATERIALS
Product Number
Brand
Product Description

Josamycin, European Pharmacopoeia (EP) Reference Standard
SKU
Pack Size
Availability
Price
Quantity
Sigma-Aldrich
Josamycin, ≥90% (HPLC)
SKU
Pack Size
Availability
Price
Quantity