GC Analysis of Multiple Pesticides on SLBยฎ-5ms after SPME using 60 ฮผm Carbowaxยฎ Fiber

CONDITIONS
sample/matrix
Pesticides at 10 ppb in 0.1M phosphate buffer, pH 7, with 25% NaCl, 9 mL in 10 mL screw cap vial
SPME fiber
60 μm Carbowax (PEG) (57354-U)
extraction
30 min. direct immersion with agitation using CombiPal autosampler
column
SLB-5ms, 30 m x 0.25 mm I.D., 0.5μm film
desorption process
2 min. at 250ยฐC, splitless for 0.75 min. then opened
oven
60ยฐC(1 min) to 150ยฐC @ 25ยฐC/min to 230ยฐC @ 5ยฐC/min to 310ยฐC @ 18ยฐC/min (3 min)
inj. temp.
Splitless/split with 0.75mm ID liner and Merlin Microseal
detector
ITMS m/z=60-450 @ 0.9 μsec per scan, transfer line 310ยฐC
carrier gas
Helium, 1mL/min constant flow