콘텐츠로 건너뛰기
Merck

GC Analysis of Organophosphorus Pesticides (OP-Pest) on SLB®-5ms after SPME using 60 μm Carbowax® Fiber

GC Analysis of Organophosphorus Pesticides (OP-Pest) on SLB®-5ms after SPME using 60 μm Carbowax® Fiber application for SPME, suitable for GC

Materials

SPME 파이버

제품 번호
설명
가격

SPME Fiber Assembly

60 μm PEG, Metal Alloy (1 cm), needle size 23 ga, Autosampler, pk of 3 ×  ea, purple hub

분석 컬럼

제품 번호
설명
가격

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.50 μm

CONDITIONS

sample/matrix

Organophosphate pesticides at 10ppb in 0.2M phasphate buffer, pH 7, with 25% NaCl, 9mL in 10 mL vial

SPME fiber

60 μm Carbowax(PEG) (57354-U)

extraction

30 min. direct immersion with agitation using CombiPal autosampler

desorption process

2 min at 250°C, splitless for 0.75 min then opened

column

SLB-5ms 30 m x 0.25mm I.D., 0.5 μm film

oven

60°C(1 min) to 215°C @ 20°C/min to 290°C @ 6°C/min (1 min)

inj. temp.

Splitless/split with 0.75mm ID liner and Merlin Microseal

detector

ITMS m/z=50-326 @ 0.6 μsec per scan, transfer line 310°C

carrier gas

Helium, 1mL/min constant flow

설명

법적 정보

Carbowax is a registered trademark of Union Carbide
SPB is a registered trademark of Merck KGaA, Darmstadt, Germany