콘텐츠로 건너뛰기
Merck

GC Analysis of Phenols on SLB®-5ms after SPME using 60 μm Carbowax® Fiber

GC Analysis of Phenols on SLB®-5ms after SPME using 60 μm Carbowax® Fiber application for SPME, suitable for GC

Materials

SPME 파이버

제품 번호
설명
가격

SPME Fiber Assembly

60 μm PEG, Metal Alloy (1 cm), needle size 23 ga, Autosampler, pk of 3 ×  ea, purple hub

분석 컬럼

제품 번호
설명
가격

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.50 μm

CONDITIONS

sample/matrix

Phenols at 25 ppb in 0.1M phosphate buffer, pH 2, with 25% NaCl, 9 mL in 10 mL screw cap vial

SPME fiber

60 μm Carbowax (PEG) (57354-U)

extraction

30 min direct immersion with agitation using CombiPal autosampler

desorption process

(30 min. direct immersion with agitation using CombiPal autosampler)

column

SLB-5ms, 30 m x 0.25 mm I.D., 0.5μm film

oven

50°C(2.0 min) to 160°C @ 10°C/min to 280°C @ 20°C/min

inj. temp.

Splitless/split with 0.75mm ID liner and Merlin Microseal

detector

ITMS m/z=45-275 @ 0.7 μsec per scan, transfer line 310°C

carrier gas

Helium, 1mL/min constant flow, 8.7 psi

설명

법적 정보

Carbowax is a registered trademark of Union Carbide
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany