GC Analysis of Triazine Herbicides on SLBยฎ-5ms after SPME using 60 ฮผm Carbowaxยฎ Fiber

GC Analysis of Triazine Herbicides on SLB®-5ms after SPME using 60 μm Carbowax® Fiber application for SPME, suitable for GC

Materials

SPME ํŒŒ์ด๋ฒ„

์ œํ’ˆ ๋ฒˆํ˜ธ
์„ค๋ช…
๊ฐ€๊ฒฉ

SPME Fiber Assembly

60 μm PEG, Metal Alloy (1 cm), needle size 23 ga, Autosampler, pk of 3 ร—  ea, purple hub

๋ถ„์„ ์ปฌ๋Ÿผ

์ œํ’ˆ ๋ฒˆํ˜ธ
์„ค๋ช…
๊ฐ€๊ฒฉ

SLBยฎ-5ms Capillary GC Column

L ร— I.D. 30 m ร— 0.25 mm, df 0.50 ฮผm

CONDITIONS

sample/matrix

Triazine herbicides at 2 ppb in 0.1M phosphate buffer, pH 7, with 25% NaCl, 9 mL in 10 mL screw cap vial

SPME fiber

60 μM Carbowax (PEG) (57354-U)

extraction

30 min direct immersion with agitation using CombiPal program

desorption process

2 min at 250ยฐC, splitless for 0.75 min then opened

column

SLB-5ms, 30 m x 0.25 mm I.D., 0.5μm film

oven

60ยฐC(1 min) to 150ยฐC @ 25ยฐC/min to 230ยฐC @ 5ยฐC/min (0.5 min)

inj. temp.

Splitless/split with 0.75mm ID liner and Merlin Microseal

detector

ITMS m/z=150-250 @ 0.5 μsec per scan, transfer line 310ยฐC

carrier gas

Helium, 1mL/min constant flow

์„ค๋ช…

๋ฒ•์  ์ •๋ณด

Carbowax is a registered trademark of Union Carbide
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany