GC Analysis of Stored Apples on 5% Phenyl Column after SPME using 100 μm PDMS Fiber

suitable for GC, application for SPME

로그인조직 및 계약 가격 보기


About This Item

기술

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

검사 파라미터

sample/matrix: 300-450g intact fruit in a 3 liter flask at ambient temp, exposed to an air flow of 25-30 mL/min
SPME fiber: polydimethylsiloxane, 100 μm film (57310-U)
desorption process: 250°C for 90 sec
column: 5% phenyl polysiloxane, 25m × 0.25 mm I.D., 0.34 μm film (Supelco equivalent, SLB-5ms, 28473-U)
oven: 40 °C (1.5 min) to 250 °C at 50 °C/min, hold 1 min
inj. temp.: 250 °C
detector: mass spectrometer, m/z = 40-300 (40 spectra/sec)
carrier gas: helium, 1.5 mL/min (44 cm/min)
sample: Apples

적합성

application for SPME

응용 분야

agriculture
food and beverages

분석 메모

SPME can reduce sampling time and is complementary to rapid separation/detection techniques such as time compressed gas chromatography (TCGC) and time-of-flight mass spectrometry (TOF-MS), to permit analysis of dozens of compounds in a few minutes. Investigators have used SPME to monitor flavor volatiles in apple, strawberry, tomato, and many other fruits and vegetables. In this application, 65 μm PDMS/divinylbenzene (DVB) SPME fibers were used to capture volatiles emanating from Mutsu apples stored for 3-4 months at 0 °C.

분석물

설명

    Hexyl hexanoate
    Isoamyl acetate
    Hexyl 2-methylbutanoate
    2-Methyl-1-butanol
    α-Farnesene
    Hexyl acetate
    Butyl acetate

가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 문서 section.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..

고객지원팀으로 연락바랍니다.